Model GFC7000E Instruction Manual

Appendix C - Repair Questionnaire - GFC7000E

048450130 Rev A1

251

Appendix C - Repair Questionnaire - GFC7000E

Company: _________________________ Contact Name: _____________________________
Phone Number: ___________ Fax Number: _____________ Email: ____________________
Site Address: __________________________________________________________________
Can we connect to the instrument? If so, provide IP address or modem #:___________________
Model GFC7000E Serial Number:_________________ Firmware

revision: _________________

The serial number can be found on the back of the instrument, the firmware revision is displayed in the upper left corner of the

display when pressing SETUP on the front panel (Example:

C.3

).

1. List all front panel error/warning messages:_________________________________________

______________________________________________________________________________

2. Please complete the following table: (Depending on options installed, not all test parameters
shown below may be available in your instrument)

PARAMETER

RECORDED

VALUE

ACCEPTABLE

VALUE

PARAMETER

RECORDED

VALUE

ACCEPTABLE

VALUE

RANGE

ppb/ppm 50 ppb - 20 ppm

SLOPE

1.0 ± 0.3

STABIL

ppb ≤ 1 ppb with zero

air

OFFSET

mV

< 250

SAMP FL

cm³/min

500 ± 50

HVPS

V

5500-900

PMT SIGNAL

WITH ZERO

AIR

mV

-20 to 150

ETEST

mV

2000 ± 1000

PMT SIGNAL

AT

SPAN GAS

CONC

mV

ppb/ppm

0-5000

0-20 000 ppb

OTEST

mV

2000 ± 1000

NORM PMT

AT

SPAN GAS

CONC

mV

ppb/ppm

0-5000

0-20 000 ppb

RCELL TEMP

°C

50 ± 1

UV LAMP

mV 2 000 to 4 000

BOX TEMP

°C

Ambient + ~5

STR. LGT

ppm ≤ 100 ppb/ zero

air

PMT TEMP

°C

7 ± 2

DARK PMT

mV

-50 to 200

IZS TEMP

°C

50 ± 3

DARK LAMP

mV

-50 to 200

3. Has the analyzer been checked for leaks? Yes No For proper flows? Yes No
4. What are the failure symptoms? _________________________________________________

______________________________________________________________________________
______________________________________________________ Continue on back if necessary

5. Which tests have you done trying to solve the problem? _______________________________

________________________________________________________

Continue on back if necessary

6. If possible, fax a portion of a strip chart or email a data file to customer service.