11.5.2. Removing/Replacing the GFC Wheel

When Removing or replacing the GFC Wheel it is important to perform the disassembly in the

following order to avoid damaging the components:

1. Turn off the analyzer.

2. Remove the top cover as described in “Getting Started” Section 3.1.

3. Open the Instrument’s hinged front panel.

4. Locate the GFC wheel/motor assembly (see Figure 3-11).

5. unplug the following electronic components:

• The GFC wheel housing temperature sensor;

• GFC wheel heater

• GFC wheel motor power supply

• IR source

6. Unscrew the purge gas line hex nut and remove the 1/8 inch FEP purge gas line.

7

8

9

10

Figure 11-7: Opening the GFC Wheel Housing

7. Remove the two (2) screws holding the opto-pickup printed circuit assembly to the GFC wheel

housing.

8. Carefully remove the opto-pickup printed circuit assembly.

9. Remove the four (4) screws holding the GFC wheel motor/heat sink assembly to the GFC

wheel housing.

10. Carefully remove the GFC Wheel motor/heat sink assembly from the GFC wheel housing.