App index – Yokogawa Removable Chassis DX1000N User Manual
Page 241
6-15
IM 04L41B01-17E
Specifications
1
2
3
4
5
6
App
Index
Register
Description
Supplementary Information
Type
Access
Simultaneous
Access
Write
Read
410611
Free message writing
Message number (1 to 10)
INT16
W
410612
Message write destination
INT16
W
• When the multi batch function is not in use
0: All groups. 1 to 36: Specified group number.
• When the multi batch function is in use
0: All groups of a specified batch number
(410613)
1 to 12: Specified group number
410613
Batch number designation for multi batch: 1 to 12
(Valid only when the multi batch function is
available. Any value when the multi batch function
is not available)
INT16
W
410614 to
410631
Free message
Up to 18 registers (up to 35 characters with '\0'
termination).
The message must be 32 characters or less.
STR36
W
410632 to
410680
(Reserved) Free
message
-
-
-
410681
Batch 1 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410682
Batch 2 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410683
Batch 3 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410684
Batch 4 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410685
Batch 5 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410686
Batch 6 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410687
Batch 7 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410688
Batch 8 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410689
Batch 9 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410690
Batch 10 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410691
Batch 11 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
410692
Batch 12 memory start
and stop
Stop: 0. Start: 1.
INT16
R/W
6.3 Modbus Protocol Specifications