App index, Lo loads setup data for setting mode, Li saves setup data – Yokogawa Removable Chassis DX1000N User Manual
Page 158: Cm sets communication input data
3-41
Commands
IM 04L41B01-17E
1
2
3
4
5
6
App
Index
p1 Mode
0
Operation mode
1
Basic setting mode
Example Set the mode to basic setting.
DS1
Description • You cannot set p1 to 1 when the DX is
recording (memory sampling) or computing, is
formatting an external storage medium, or is
storing data to an external storage medium.
• You cannot set p1 to zero when the DX is
formatting an external storage medium or is
storing data to an external storage medium.
• To activate the settings you have changed
using basic setting commands, you must use
the XE command to save the settings. Be sure
to use the XE command to save the settings
before switching the execution mode back to
operation. If you do not save the settings and
change the execution mode back to operation,
the DX returns to the previous settings.
• This command is invalid on models with the
/AS1 advanced security option.
LO
Loads setup data for setting
mode
Syntax
LO p1,p2
p1 File name (up to 32 characters)
p2 Medium
0
CF slot
1
USB
Example Load setup data for setting mode from the setup
file named SETFILE1.
LOSETFILE1
Description • Do not specify the extension when specifying
the file name.
• You can set p2 to 1 on models with the /USB1
USB interface option.
• If you omit parameter p2, the medium is set to
CF slot.
• You cannot use this command to load setup
data for basic setting mode. To load setup
data for both setting and basic setting modes,
use the YO command.
• You cannot use this command when there is
no external storage medium inserted in the
DX.
LI
Saves setup data
Syntax
LI p1
p1 File name (up to 32 characters)
p2 Medium
0
CF slot
1
USB
Example Saves setup data for both setting and basic
setting modes to a file named SETFILE2 on the
CF card.
LISETFILE2
Description • Do not specify the extension when specifying
the file name.
• You can set p2 to 1 on models with the /USB1
USB interface option.
• If you omit parameter p2, the medium is set to
CF slot.
• An extension is attached to the file that you
save.
• You cannot use this command when there is
no external storage medium inserted in the
DX.
CM
Sets communication input data
Syntax
CM p1,p2
p1 Communication input datal number
p2 Communication input data
The selectable range is -9.9999E+29
to -1.0000E-30, 0, and 1.0000E-30 to
9.9999E+29.
Five significant digits
Query
CM?
Example Enter 1.0000E-10 to communication input data
C01.
CMC01,1.0000E-10
Description • You can use this command on models with the
/M1 or /PM1 option.
• On models with the /CP1 PROFIBUS-DP
interface option, the communication input
data for C01 to C24 (on the DX1000) or for
C01 to C32 (on the DX2000) is reserved for
PROFIBUS-DP. The client device cannot
specify values for this communication input
data.
CE
Sets communication input of an
external input channel
Syntax
CE p1,p2
p1 External input channel number
p2 Data value (-30000 to 30000)
Query
CE[p1]?
Example Set external input channel number 440 to 12345.
CE440,12345
Description You can use this command on models with the
/MC1 external input channel option.
EM
Starts or stops the e-mail
transmission function
Syntax
EM p1
p1 Type of operation
0
Start
1
Stop
3.5 Control Commands