Measured/computed data and fifo data, Number of blocks, Number of bytes – Yokogawa Removable Chassis DX1000N User Manual

Page 218: Block, Flag

background image

4-34

IM 04L41B01-17E

Measured/Computed Data and FIFO Data

• The FD command is used to output the measured/computed data.
• The FF command is used to output the FIFO data.
• You can use the CB command to specify whether to output the data of measurement

channels set to skip and computation or external input channels set to OFF.

• The ID number of the output format is 1. See “ID” in section 4.1.

2 byte

Number of blocks

Block 1

Block n

Number of bytes

2 byte

Binary data

(The binary data section on the

“Conceptual diagram” in

section 4.1.)

Number of Blocks

This is the number of blocks.

Number of Bytes

This is the size of one block in bytes.

Block

Type

Type

Type

Year

Month

Channel

Channel

Channel

Day

A2A1

A2A1

A2A1

Hour

A4A3

A4A3

A4A3

Min

ms

Summer/

winter

Flag

s

Measured data

External input data

Computed data

1 byte 1 byte 1 byte 1 byte 1 byte

1 byte

2 bytes

1 byte

1 byte

4 bits

12 bits

• Flag

The meaning of the each flag is given in the table below. The flags are valid during

FIFO data output. The flags are undefined for other cases.

Bit

Flag

Meaning of the Flag

0

1

7

No

Yes

Indicates that the screen snapshot was executed.

6

5

4

3

2

No

Yes

Indicates that the decimal position or unit information was

changed during measurement.

1

No

Yes

Indicates that the FIFO acquiring interval was changed with

the FR command during measurement.

0

No

Yes

Indicates that the internal process took too much time

(computation, for example) and that the measurement could

not keep up at the specified scan interval.

The bits that have “•” for the flag column are not used. The value is undefined.