App index, Xb sets burnout detection, Xj sets rjc – Yokogawa Removable Chassis DX1000N User Manual
Page 166: Xt sets the temperature unit, Rf sets key lock
3-49
Commands
IM 04L41B01-17E
1
2
3
4
5
6
App
Index
p3 Scan interval (25MS, 125MS, 250MS, 1S,
2S, 5S)
p4 A/D integration time (AUTO, 600Hz, 50Hz,
60Hz, 100ms)
Query
XV?
Example Set the scan interval to 1 second in normal
mode.
XV1,NORMAL,1S
Description • The combinations of available scan interval
modes and scan intervals vary depending
on the model. For details, see the DX1000/
DX1000N or DX2000 User’s Manual
(IM04L41B01-01E or IM04L42B01-01E).
• You can set p4 to 600 Hz for fast sampling
mode. You can choose 100 ms when the scan
interval is set to 2 s or 5 s.
• On models with multi batch /BT2, you can only
set p2 to NORMAL and p3 to 1S, 2S, or 5S.
XB
Sets burnout detection
Syntax
XB p1,p2
p1 Measurement channel number
p2 Burnout processing
OFF
No processing
UP
Sets the computed result to
positive overflow.
DOWN
Sets the computed result to
negative overflow.
Query
XB[p1]?
Example Set the measured result to UP (positive overflow)
when channel 001 burns out.
XB001,UP
Description Set p1 by referring to the table in section 3.3.
XJ
Sets RJC
When using the internal compensation circuit
Syntax
XJ p1,p2
p1 Measurement channel number
p2 RJC mode (INTERNAL)
Query
XJ[p1]?
Example Set the channel 001 RJC to internal
compensation circuit.
XJ001,INTERNAL
When using an external RJC
Syntax
XJ p1,p2,p3
p1 Measurement channel number
p2 RJC mode (EXTERNAL)
p3 External RJC value (-20000 to 20000)
Query
XJ[p1]?
Example Set the channel 002 RJC to external, and set the
compensation value to 0 μV.
XJ002,EXTERNAL,0
Description • Set p1 by referring to the table in section 3.3.
• The unit of p3 is the μV.
XM
Sets memory sampling
conditions
Syntax
XM p1
p1 Data type
DISPLAY Display data
EVENT Event data
E+D
Display data and event data
Query
XM?
Example Set the memory sampling condition to display
data.
XMDISPLAY
Description You cannot specify E+D when:
• Multi batch /BT2 is in use.
• Trend interval switching is on.
• You are using a DX with the /AS1 advanced
security option.
XT
Sets the temperature unit
Syntax XT p1
p1 Temperature unit (C, F)
Query
XT?
Example Set the temperature unit to Celsius.
XTC
RF
Sets key lock
When p1 is set to KEY
Syntax
RF p1,p2,p3,p4,p5,p6,p7
p1 Type (KEY)
p2 START key (FREE, LOCK)
p3 STOP key (FREE, LOCK)
p4 MENU key (FREE, LOCK)
p5 USER key (FREE, LOCK)
p6 DISP/ENTER key (FREE, LOCK)
p7 FAVORITE key (FREE, LOCK)
When p1 set to FUNC (function key)
Syntax
RF p1,p2,p3,p4,p5,p6,p7,p8
p1 Type (FUNC)
p2 Alarm ACK (FREE, LOCK)
p3 Message/batch key (FREE, LOCK)
p4 Math key (FREE, LOCK)
p5 Data save (FREE, LOCK)
p6 E-mail/FTP (FREE, LOCK)
p7 Time set (FREE, LOCK)
p8 Display Function (FREE, LOCK)
When p1 is set to MEDIA (external storage media)
Syntax
RF p1,p2,p3
p1 Type (MEDIA)
p2 External storage media operation (FREE,
LOCK)
p3 Setup loading operation (FREE, LOCK)
Query
RF[p1]?
Example Lock the MENU key (leave other keys unlocked).
RFKEY,FREE,FREE,LOCK,FREE,FREE,FREE