9 output commands (rs-422/485 dedicated commands), Esc o opens an instrument, Esc c closes an instrument – Yokogawa Removable Chassis DX1000N User Manual
Page 181
3-64
IM 04L41B01-17E
p2 Output data type
MANUAL Manual sampled data
REPORT Report
p3 Specified file name
Example Output report data,
000142_080102_004127
H_.DAR from the DX.
MOGET,REPORT,000142_080102_004127H_.
DAR
Description Parameter p3 is valid when p1 is set to GET or
SIZE.
3.9 Output Commands
(RS-422/485 Dedicated
Commands)
ESC O Opens an instrument
ESC in ASCII code is 1BH. For details, see
appendix 3.
Syntax
ESC O p1
p1 Instrument address (01 to 99)
Example Open the instrument at address 99, and enable
all commands.
ESC O99
Description • Specifies the address of the instrument that
you want to communicate with.
• You can only open one instrument at any
given time.
• If you execute ESC O, any instrument that is
already open is automatically closed.
• When the DX receives this command
successfully, the DX returns “
ESC O
”.
• Normally, the terminator can be CR+LF or LF
for communication commands. However, you
must terminate this command with CR+LF.
ESC C Closes an instrument
ESC in ASCII code is 1BH. For details, see
appendix 3.
Syntax
ESC C p1
p1 Instrument address (01 to 99)
Example Close the device whose address is 77.
ESC C77
Description • This command closes the connection to the
instrument you are communicating with.
• When the DX receives this command
successfully, the DX returns “
ESC C
”.
• Normally, the terminator can be CR+LF or LF
for communication commands. However, you
must terminate this command with CR+LF.
3.8 Output Commands (Setting,...) / 3.9 Output Commands (RS-422/485 ...)