7 output commands (control), 7 output commands (control) -60, Bo sets the output byte order – Yokogawa Removable Chassis DX1000N User Manual
Page 177: If sets status filters, Cb sets the data output format
3-60
IM 04L41B01-17E
3.7 Output Commands
(Control)
BO
Sets the output byte order
Syntax
BO p1
p1 Byte order
0
Outputs data MSB first.
1
Outputs data LSB first.
Query
BO?
Example Output data MSB first.
BO0
Description This command applies to the byte order of
numeric data for BINARY output.
CS
Sets the check sum (can
only be used during serial
communications)
Syntax
CS p1
p1 Checksum usage
0
Do not calculate (value fixed at
zero)
1
Calculate
Query
CS?
Example Enable (Calculate) the checksum.
CS1
Description You can use this command only for serial
communications.
IF
Sets status filters
Syntax
IF p1,P2
p1 Filter values for status information numbers
1 to 4
(0.0.0.0 to 255.255.255.255)
p2 Filter values for status information numbers
5 to 8
(0.0.0.0 to 255.255.255.255)
Query
IF?
Example Set the status filter values to 1.0.4.0 and
255.127.63.31.
IF 1.0.4.0,255.127.63.31
Description For details, see chapter 5.
CB
Sets the data output format
Syntax
CB p1
p1 Output format
0
Normal output (includes data from
channels set to SKIP and OFF)
1
Do not output data from channels
set to SKIP or OFF
Query
CB?
Example Set the output format to normal output.
CB0
Description • This setting is separate for each connection.
• This command only affects the communication
section and does not affect the front panel
settings.
• Effective range of commands
Output information
Corresponding
command
Instantaneous data output (binary)
FD1, FF
Instantaneous data output (ASCII)
FD0
Decimal place information (ASCII)
FE1
Setup channel information (binary)
FE5
Configured alarm information (binary) FE6
CC
Disconnects the Ethernet
connection (can only be used for
Ethernet communications)
Syntax
CC p1
p1 Disconnection (0)
Example Disconnect the connection.
CC0
Note
Initialization of settings specified using the BO, CS, IF, and
CB commands
• Serial communications
Settings specified using the BO, CS, IF, and CB commands
are reset to the following default values when you reset the DX
(when you turn the DX off and then back on or when you exit
from basic setting mode).
• Output byte order, checksum, output format: 0
• Status filter: 255.255.255.255
If you reset the DX, you must set these values again.
• Ethernet communications
Settings specified using the BO, IF, and CB commands
are reset to their default values when you disconnect the
connection to the DX. After reconnecting to the DX, set these
values again.
3.7 Output Commands (Control)