beautypg.com

Communication log – Yokogawa Removable Chassis DX1000N User Manual

Page 194

background image

4-10

IM 04L41B01-17E

Communication Log

• The FL command is used to output the data.
• A log of setting/basic setting/output commands and responses is output. Up to 200

logs are retained. Logs that exceed 200 are cleared from the oldest data.

• Syntax

EACRLF
yy/mo/dd_hh:mm:ss_n_uuu···ufd_mmm···mCRLF
······································
ENCRLF

yy Year (00 to 99)
mo Month (01 to 12)
dd Day (01 to 31)
hh Hour (00 to 23)
mm Minute (00 to 59)
ss Second (00 to 59)
n Connection ID. A number used to identify the user that is connected.

0: Serial
1 to 3: Ethernet

uuu···u User name (up to 20 characters)
f Multiple command flag
Space: Single
*: Multiple
(If multiple commands are separated by sub delimiters and output at

once, “*” is displayed. The multiple commands are divided at each sub
delimiter and stored as individual logs (1 log for 1 command and 1 log
for 1 response.)

d Input/Output

>: Input
<: Output

mmm···m Message (up to 20 characters)

• The communication log contains only the error number and not the

error message section.

• Normally, the transfer data are transmitted as they are, but in some

cases, a special message is output. The special messages are
shown below.

Reception

(Over length):

Command length exceeded.

(Over number):

Number of commands exceeded.

(Serial error): Received an error character through serial

communications.

4.2 Output Format of ASCII Data

This manual is related to the following products: