ddddd Mantissa (00000 to 99999, 5 digits)
• Eight digits for computed data.
• For abnormal data (data status is E) or data of which the mantissa or
the exponent exceeds the range (data status is O), the mantissa is
set to 99999 (99999999 for computed data).
pp Exponent (00 to 04)
_ Space
•
Example
EA
DATE 99/02/23
TIME 19:56:32.500
N 001h mV +12345E-03
N 002 mV -67890E-01
S 003
EN
Note
• Data for non-existing channels are not output (not even the channel number).
• For channels set to skip, output values from alarm status to exponent are spaces.
Relay Status and Internal Switch Status
The FD command is used to output the DO status and internal switch status.
• Syntax
EACRLF
I01-I06:aaaaaaCRLF
I11-I16:aaaaaaCRLF
I21-I26:aaaaaaCRLF
I31-I36:aaaaaaCRLF
S01-S30:aaa···CRLF
ENCRLF
aaa··· Indicates the relay statuses in ascending order by relay number from the
left.
1: Relay ON
0: Relay OFF
-: Relay not installed
• Example 1
When relays I01 to I04 are ON, and I05 and I06 are not installed (for the DX1000).
EA
I01-I06:1111--
I11-I16:------
I21-I26:------
I31-I36:------
S01-I30:000000000000000000000000000000
EN