beautypg.com

Chapter 6: sensor calibration, When to calibrate sensor, Chapter 6 – RED DIGITAL CINEMA DSMC2 BRAIN with HELIUM 8K S35 Sensor (2018 Unified DSMC2 Lineup) User Manual

Page 159: Sensor calibration, Dsmc2 helium operation guide

background image

C O PYR I G HT © 2 0 1 9 R ED.C O M , LLC

9 5 5 - 0 1 6 8 _V 7 .1 , R EV - M    

|

1 5 9

DSMC2 HELIUM OPERATION GUIDE

CHAPTER 6:

SENSOR CALIBRATION

Sensor calibration, also known as black shading, is a process during which the camera optimizes image quality by
ensuring that pixel sensitivity remains consistent throughout the sensor. Calibrating the sensor generates a calibration
map based on system and environment settings. After calibration, the sensor shows a uniform noise profile over the
whole area without any falloffs on any side.

The camera offers two (2) ways to calibrate the sensor:

Manual Calibration

: RED recommends using Manual Calibration for the HELIUM 8K S35, MONSTRO 8K VV, and

GEMINI 5K S35 sensors. For more information, go to

"Calibrate Sensor: Manual Calibration" on the next page

.

Auto Calibration

: RED recommends using Auto Calibration for the DRAGON 4.5K, DRAGON 5K, DRAGON 6K,

and DRAGON-X 5K sensors. For more information, go to

"Calibrate Sensor: Auto Calibration" on page 161

.

For more information, see the

Sensor Calibration Methods

article.

NOTE:

Calibration maps are still available after performing a system restore.

WHEN TO CALIBRATE SENSOR

Calibration is required:

After an extreme change in temperature (+/– 30°F or +/–15°C) from the current calibration map.

After an extreme change in exposure time (+/– 1/2 sec) from the current calibration map.

If either the T or E in the CAL: T/E indicator is not green.

After each firmware upgrade.

To check the temperature/exposure of the current calibration file, go to

Menu

>

Settings

>

Maintenance

>

System

Status

>

Project Status

.