PEL-3000 User Manual

96

8. Choose the P/F-Time (pass/fail time) in

seconds.

The P/F-Time refers to the P/F delay time. This
delay time includes the 0.06 P/F start test time,
as shown in the timing diagram on page 94.

P/F-Time

Off, 0.0 ~ 119.9 seconds

9. Set the Short-Time in seconds.

Has the same action as pressing the short key.
See page 61 for details about shorting the load.

Short-Time

Off, 0.1 seconds ~ On-Time

10. Repeat steps 3 to 9 for all the steps in the

program.

A maximum of 16 steps per program can be
created.

Steps that are not configured are set to ―Skip‖
by default.

11. Press Save[F3] to save the program and all the

steps in the program.

The program will be saved to internal memory.

See the Save/Recall chapter on details on how
to save to Setup memory.

Recall Default

Pressing Recall Default[F4] will recall the default
settings for each program/step. See page 178
for details.