EXTERNAL CONTROL

153

Parallel Capacity, PEL-3211

Model No. of

Units

V

I

Total Sink Current

PEL-3111 +

PEL-3211

Total Power

PEL-3111 +

PEL-3211

PEL-3111:

Master

x 1 150V 210A

N/A

N/A

PEL-3211:

Slave

Boosters

x 1 150V 420A

630A

3150W

x 2 150V 840A

1050A

5250W

x 3 150V 1260A

1470A

7350W

x 4 150V 1680A

1890A

9450W

Note: The PEL-3211 booster packs do not have a control panel. They can
only be used as slaves with a single PEL-3111 in parallel.

Connection

Description

The J1 and J2 connectors are used for control
during parallel operation. Up to 5 units can be
used in parallel.

Note

Only the rear panel terminals can be used for
parallel operation, the front panel terminals have a
lower current rating and thus should not be used
for parallel operation.