OPERATION

73

Von Voltage Delay

Description

Von Delay is the amount of time the unit will
wait before turning the load on after the Von
Voltage threshold has been latched. This will
prevent overshoot current from affecting the
Von Voltage threshold.

Operation

2. Press

Main

> Configure[F5] > Other[F2] and

set the Von Delay time.

Range:

Von Delay: OFF, 1.0~60ms


Note: CR mode can have the delay time set
separately from the other modes (called

Von Delay

–CR

when in CR mode).

Timer Functions

Count Time

Description

When Count Time is set to on, it will count the
elapsed time from when the load was turned on
to when it was turned off.

This function is applicable to manual and
automatic shutdown (such as from protection
functions such as UVP etc.)

The elapsed time will be shown in the display
Measurement area.