PEL-3000 User Manual

64

Level

Time

Level 2

Level 1

Timer1 Timer2 Timer1

Dynamic Mode

Note

Dynamic mode is not available for CV mode.

Operation

1. Make sure the load is off.

2. Press

Main

.

3. Select Dynamic or Static mode with the

Function[F4] soft-key.

A different switching mode can be set for CC,
CR and CP mode.

4. For dynamic mode, set the Timer1 and Timer2

parameters using the scroll wheel and number
pad.

Timer1 sets the Level1 on-time.

Timer2 sets the Level2 on-time.

Take the slew rate settings into consideration
when setting the timers.

The frequency of the dynamic switching is
output via the TRIG OUT BNC.

To select whether A Value or B Value is the
―active‖ setting, press the

Shift

+

Preset

keys.

The ―active‖ value will be shown in the
Operation Status Panel.