OPERATION

83

Display

2.596

V

500.06

mA

1.2982

w

CV B Value

15.000

CV A Value

15.000

LOAD

RS232

A Value

Fine

03/Sep/2012

V

V

UVP Alarm

Please Press Enter

To Clear Alarm

UVP

UVP indicator

Alarm message
when UVP is
tripped

Mode

CV

I Range

L 0.7A

V Range

L 15V

Response

Fast

Configure

OVP

Description

If the OVP is tripped, the PEL-3000 will turn off
the load.

The OVP levels can be set from 0V to 10%
higher than the rating voltage.

Operation

1. Press

Main

> Configure[F5] > Protection[F1]

and set the OVP Level.

Range:

OVP Level: OFF, 0~Rating voltage +
10%

Note: To turn OVP off, set the OVP voltage greater
than the current rating voltage + 10%.

Alarm

The OVP indicator and a message will only
appear on the screen when the input voltage is
below the UVP level. The Enter key must be
pressed to clear the alarm message.

To clear the OVP indicator, remove the cause of
the over voltage ~ i.e., reduce the input voltage.