78

Step Mode Configuration

Description

When set to Step Mode, the voltage, current,
resistance and power settings can have the step
resolution configured. The step resolution
refers to the step resolution of the coarse
adjustment for these settings. The fine
adjustment cannot be configured.

See the Conventions section on page 41 for
details on how to switch between coarse and
fine adjustment modes.

Settings

The step resolution of each setting is configured
separately for each current range.

Settings

Description

CCH Step

CC mode, IRange = High

CCM Step

CC mode, IRange = Middle

CCL Step

CC mode, IRange = Low

CRH Step

CR mode, IRange = High

CRM Step

CR mode, IRange = Middle

CRL Step

CR mode, IRange = Low

CVH Step

CV mode, VRange = High

CVL Step

CV mode, VRange = Low

CPH Step

CP mode, IRange = High

CPM Step

CP mode, IRange = Middle

CPL Step

CP mode, IRange = Low