176

Replace the Clock Battery

Background

The system clock keeps time using a user-
replaceable battery.

The battery should be replaced approximately
every 3 years.

Battery type:

CR123A

Procedure

1. Turn off the PEL-3000 and remove the case.

First remove the handle by carefully removing
the plastic tabs and then unscrewing the two
screws connecting the handle to the case.

A total of 10 screws should be removed from the
case.

2. Remove the battery and replace with the same

type and rating.

The battery is located on the right hand side,
near the rear panel.

CR123A