Dynamic mode – GW Instek PEL-3000 Series User Manual User Manual
Page 202
PEL-3000 User Manual
200
Ammeter
H, M Range 0.000A~35.000A
0.000A~70.000A
0.00A~210.00A
L Range
0.00mA~350.00mA
0.00mA~700.00mA
0.0mA~2100.0mA
Accuracy
±(0.2 % of rdg + 0.3 % of f.s)
Accuracy
Parallel Operation: ±(1.2% of rdg +1.1% of f.s.)
Wattmeter
H, M Range 0.00W~175.00W
0.00W~350.00W
0.00W~1050W
L(CC/CR/CV
mode)
0.000W~52.500W
0.000W~ 105.000W
0.00W~315.00W
L(CP mode) 0.0000W~ 1.7500W
0.0000W~ 3.5000W
0.000W~ 10.500W
Temperature Coefficient per ºC
Voltmeter
100ppm
Ammeter
200ppm
Dynamic Mode
Model
PEL-3021
PEL-3041
PEL-3111
Operating Mode
CC, CR and CP
T1 & T2
0.025ms ~ 10ms / Res: 1us
10ms ~ 30s / Res: 1ms
Accuracy
1us / 1ms ± 100ppm
Frequency Range (Freq./Duty)
1Hz ~20kHz
Frequency Resolution
1Hz~9.9Hz
0.1Hz
10Hz~99Hz
1Hz
100Hz~990Hz
10Hz
1kHz~20kHz
100Hz
Frequency Accuracy of Setting
(0.5% of set)
Duty Cycle of Setting (Freq./Duty)
1% ~99% , 0.1% step
The minimum time width is 10 us. Between 1kHz and 20kHz, the
maximum duty cycle is limited by the minimum time width.
Slew Rate Setting Range (CC Mode)
H Range
2.5mA/us~2.5A/us
5mA/us~5A/us
16mA/us~16A/us
M Range
250uA/us~250mA/us 500uA/us~500mA/us 1.6mA/us~1.6A/us
L Range
25uA/us~25mA/us
50uA/us~50mA/us
160uA/us~160mA/us