Parallel Operation

The PEL-3000 series can be connected in parallel to increase the
total power capacity of a single unit.
The PEL-3000 series can operate with up to 5 units in parallel. A
single unit is designated as a master unit and any other connected
units as slaves.
Only units of the same type and rating can be used in parallel or
alternatively, the PEL-3211 booster pack can be used as a slave with
the PEL-3111.
When a master unit is used in parallel mode, to ensure stability, the
response speed will drop down to 1/2 if it was originally 1/1. You
can however, reset the response speed back (or to another value) in
the Main>Configure menu.

Parallel Capacity, PEL-3021, PEL-3041, PEL-3111

Model

Single Unit 2 Units

3 Units

4 Units

5 Units

PEL-3021

150V

35A

175W

150V

70A

350W

150V

105A

525W

150V

140A

700W

150V

175A

875W

PEL-3041

150V

70A

350W

150V

140A

700W

150V

210A

1050W

150V

280A

1400W

150V

350A

1750W

PEL-3111

150V

210A

1050W

150V

420A

2100W

150V

630A

3150W

150V

1680A

4200W

150V

1050A

5250W