Frame control connector contacts – GW Instek PEL-3000 Series User Manual User Manual
Page 183

APPENDIX
181
Frame Control Connector Contacts
J1 Connector
Pin name
Pin number
Description
EXT R/V CONT
1 Used for voltage/resistance control of CC, CR, CV
and CP mode. 
 
0V to 10V corresponds to 0% to 100% of the rated 
current (CC mode), rated voltage (CV mode), or 
rated power (CP mode). 0V to 10V corresponds to 
the maximum resistance to minimum resistance 
(CR mode) 
 
0Ω to 10kΩ corresponds to 0% to 100% or 100% 
to 0% of the rated current (CC mode), rated 
voltage (CV mode), or rated power (CP mode). 0Ω 
to 10kΩ corresponds to maximum resistance to 
minimum resistance or minimum resistance to 
maximum resistance (CR mode) 
IMON
2 Current monitor output
10 V f.s (H/L range) and 1 V f.s (M range)
A COM
3 Connected to the negative load input terminal on
the rear panel.
SUM I MON
4 Used during master/slave operation. Connected to
SUM I MON of the J2 connector.
PRL IN+
5 Used during master/slave operation. Connected to
PRL OUT+ of the J2 connector.
PRL IN-
6 Used during master/slave operation. Connected to
PRL OUT- of the J2 connector.
LOAD ON/OFF 
CONT 
7 Turns on the load with low (or high) TTL level
signal 
Pulled up the internal circuit to 5 V using 10 kΩ. 
RANGE CONT 1 8 External range switch input*1 *2
Pulled up the internal circuit to 5 V using 10 kΩ.
RANGE CONT 0 9 
ALARM INPUT 
10 Activates alarm with low TTL level signal input.
Pulled up the internal circuit to 5 V using 10 kΩ.