APPENDIX

181

Frame Control Connector Contacts

J1 Connector

Pin name

Pin number

Description

EXT R/V CONT

1 Used for voltage/resistance control of CC, CR, CV

and CP mode.

0V to 10V corresponds to 0% to 100% of the rated
current (CC mode), rated voltage (CV mode), or
rated power (CP mode). 0V to 10V corresponds to
the maximum resistance to minimum resistance
(CR mode)

0Ω to 10kΩ corresponds to 0% to 100% or 100%
to 0% of the rated current (CC mode), rated
voltage (CV mode), or rated power (CP mode). 0Ω
to 10kΩ corresponds to maximum resistance to
minimum resistance or minimum resistance to
maximum resistance (CR mode)

IMON

2 Current monitor output

10 V f.s (H/L range) and 1 V f.s (M range)

A COM

3 Connected to the negative load input terminal on

the rear panel.

SUM I MON

4 Used during master/slave operation. Connected to

SUM I MON of the J2 connector.

PRL IN+

5 Used during master/slave operation. Connected to

PRL OUT+ of the J2 connector.

PRL IN-

6 Used during master/slave operation. Connected to

PRL OUT- of the J2 connector.

LOAD ON/OFF
CONT

7 Turns on the load with low (or high) TTL level

signal
Pulled up the internal circuit to 5 V using 10 kΩ.

RANGE CONT 1 8 External range switch input*1 *2

Pulled up the internal circuit to 5 V using 10 kΩ.

RANGE CONT 0 9
ALARM INPUT

10 Activates alarm with low TTL level signal input.

Pulled up the internal circuit to 5 V using 10 kΩ.