Page 140

background image

PEL-3000 User Manual

138

3. Turn the power on the PEL-3000.

4. Set the operating mode and range.

See page 51 for CC mode.

See page 53 for CR mode.

See page 55 for CV mode.

See page 56 for CP mode.

5. Press

Main

> Configure [F5] > Next Menu [F4]

> External [F3].

6. Set the Control parameter to V.

The J1 connector is now ready for external
voltage control.

External Resistance Control - Overview

Background

External resistance control of the CC, CR, CV
and CP modes is accomplished using the J1
connector on the rear panel.

A resistance of 0kΩ~10kΩ is used to control the
input current, voltage, resistance or power on
the PEL-3000.

The input can be configured to vary in
proportion to the external resistance or the
inverse. See page 140 for more details on
proportional and inverse resistance control.

Note

Exceeding 11.8k

Ω will cause an EXT.OV alarm

message which will reset the voltage output to 0
until the external resistance is reduced back down
below 11.8k

Ω.