Cv mode – GW Instek PEL-3000 Series User Manual User Manual

Page 57

background image

OPERATION

55

CR Units

Description

The CR setting units can be set to ohm (Ω) or
millisiemens (mS).

Operation

1. Make sure the load is off.

2. Press

Main

> Configure[F5] > Other[F2] and

set the CR Unit setting.

Range:

Ω, mS

CV Mode

Description

In Constant Voltage Mode, the unit will
maintain a constant voltage. In CV mode you
set the constant voltage level. For more details
on CV mode, see the appendix on page 190.

Warning

If you change the mode or the range when the load
is already on, the load will be turned off
automatically.

Operation

1. Make sure the load is off.

2. Press

Main

.

3. Select CV mode with the Mode[F1] soft-key.

4. Select the current range with the I Range[F2]

soft-key.

Range:

High, Middle, Low

5. Select the voltage range with the V Range[F3]

soft-key.

Range:

High, Low