Protection Function

Model

PEL-3021

PEL-3041

PEL-3111

Overvoltage protection(OVP)

Turns off the load at 110% of the rated voltage

Overcurrent protection(OCP)

0.03 ~ 38.5A

0.06A ~ 77A

0.2A ~ 231A

or 110% of the maximum current of each range
Load off or limit selectable

Overpower protection(OPP)

0.1W ~ 192.5W

0.3W ~ 385W

1W ~ 1155W

or 110% of the maximum power of each range
Load off or limit selectable

Overheat protection(OHP)

Turns off the load when the heat sink temperature reaches 95 °C

Undervoltage protection(UVP)

Turns off the load when detected. Can be set in the range of 0 V to
150 V or Off.

Reverse voltage protection(RVP)

By diode. Turns off the load when an alarm occurs.

Rating overcurrent protection (R.OCP)

An R.OCP message will be produced when the input current range is
greater than 110% of the rated operating current range (I range).

Sequence

Normal Sequence
Operation
mode

CC, CR, CV or CP

Maximum
number of
steps

1000

Step
Execution
Time

1ms – 999 h 59 min

Time
resolution

1 ms (1 ms – 1 min)/100 ms (1 min – 1 h)/1 s (1 h – 10 h)/10 s (10
h – 100 h)/1 min (100 h – 999 h 59 min)