GW Instek PEL-3000 Series User Manual User Manual

Page 76

background image

PEL-3000 User Manual

74

Operation

1. Press

Main

> Configure[F5] > Other[F2] and

turn the Count Time on or off.

Range:

ON, OFF

Display

Mode

CP

IRange

H 35A

VRange

L 15V

Function

Dynamic

Configure

0.000

V

0.000

A

0.00

w

Timer1

0.025

Level2

0.00

Level1

0.00

LOAD

RS232

Fine

03/Sep/2012

mS

W

W

0:00:05

Elapsed time

Cut Off Time

Description

The Cut Off Time function will turn the load off
after a set-amount of time. After the load has
been turned off, a popup screen will display the
voltage level when the load was turned off.

Operation

1. Press

Main

> Configure[F5] > Other[F2] and

set the Cut Off Time.

Range:

OFF, 1 second ~ 999 hours:59
minutes:59 seconds

Display

Mode

CP

I Range

H 35A

V Range

L 15V

Function

Dynamic

Configure

0.000

V

0.000

A

0.00

w

Timer1

0.025

Level2

0.00

Level1

0.00

LOAD

RS232

Fine

03/Sep/2012

mS

W

W

0:00:05

Cut off time

Time Up

Voltage : 5.1223V

Enter

Voltage at cut off time