Advanced Configuration Settings

Use the advanced configuration settings to configure settings other
than those described in the basic configuration chapter.

Soft Start Setting

Description

The soft start setting is used to limit the amount
of input current at start-up or from when the
Von Voltage threshold is tripped.

The soft start setting only applies to CC, CR and
CP mode.

Input I

Time

Soft Start = ON

Input I

Time

Soft Start= OFF

Rise voltage

Rise current

Rise current

Rise voltage

Operation

1. Press

Main

> Configure[F5] > Other[F2] and

set the Soft Start time.

Range:

OFF, 1~200ms