89

Go-NoGo

The Go-NoGo configuration is used to create pass/fail limits on the
voltage or current input. If the voltage/current exceeds the
pass/fail limits, an alarm will be output.

The Go-NoGo configuration can be used with the Program function
to create complex pass/fail tests.

Setting the Go-NoGo Limits

Description

The Go-NoGo setting limits can be set as either
discrete high & low values or as a percentage
offset from a center value.

Operation

1. Press

Main

> Configure[F5] > Go-NoGo[F3].

2. Select Entry Mode and choose how to set the

pass/fail limits.

Value will allow you to set the limits as discrete
values.

Percent will allow you to set the limits as a
percentage offset from a center value.

3. If Entry Mode was set to Value, Set the High &

Low limit values.

High:
Low:

0~rating current/voltage
0 ~rating current/voltage

4. If Entry Mode was set to Percent, Set the Center

voltage/current and High, Low % values.

Center:
High:
Low:

0~rating current/voltage
0~100% of center voltage/current
0~100% of center voltage/current