beautypg.com

Cc mode – GW Instek PEL-3000 Series User Manual User Manual

Page 198

background image

PEL-3000 User Manual

196

CC Mode

Model

PEL-3021

PEL-3041

PEL-3111

Operating Range
H Range

0A~35A

0A~70A

0A~210A

M Range

0A~3.5A

0A~7A

0A~21A

L Range

0A~0.35A

0A~0.7A

0A~2.1A

Setting Range
H Range

0A~35.7A

0A~71.4A

0A~214.2A

M Range

0A~3.57A

0A~7.14A

0A~21.42A

L Range

0A~0.357A

0A~0.714A

0A~2.142A

Default Setting
H Range

0A

0A

0A

M Range

0A

0A

0A

L Range

0A

0A

0A

Resolution
H Range

1mA

2mA

10mA

M Range

0.1mA

0.2mA

1mA

L Range

0.01mA

0.02mA

0.1mA

Accuracy of Setting
H, M Range ±(0.2 % of set + 0.1 % of f.s

*1

) + Vin

*2

/500 kΩ

L Range

±(0.2 % of set + 0.1 % of f.s) + Vin

*2

/500 kΩ

Parallel
Operation

±(1.2% of set +1.1% of f.s

.*3

)

Input Voltage Variation

*4

H Range

2mA

4mA

10mA

M Range

2mA

4mA

10mA

L Range

0.1mA

0.2mA

0.6mA

Ripple
RMS

*5

3mA

5mA

20mA

*7

P-P

*6

30mA

50mA

100mA

*7


*1 Full scale of H range
*2 Vin: input terminal voltage of electronic load
*3 M range applies to the full scale of H range
*4 When the input voltage is varied from 1.5V to 150V at a current of rated

power/150V

*5 Measurement frequency bandwidth: 10Hz to 1MHz
*6 Measurement frequency bandwidth: 10Hz to 20MHz
*7 At measurement current of 100A