OPERATION

63

Basic Configuration

The basic configuration settings are the common configuration
settings that are used for each operating mode. After selecting a
basic operating mode (CC, CR, CV or CP mode), the slew rate,
switching mode, response rate and other common parameters
should be configured.

Select the Switching Function

Description

The PEL-3000 has two switching modes, static
and dynamic. The switching modes allow the
PEL-3000 to switch between two preset levels.
Static mode can only switch between the two
levels manually, while Dynamic mode switches
between each level automatically based on a
timer.

Static mode: A Value, B Value

Dynamic mode: Level1, Level2

When the unit is set to static mode, only one
value (A Value or B Value) can be active at a
time. The active value is shown in the
Operation Status Panel.

Function

Static

Configure

0.00

w

LOAD

RS232

SHORT

A Value

Fine

W

W

Active setting

When the unit is set to dynamic mode, the unit
will switch between Level1 and Level2 based
on the Timer1 and Timer2 parameters, shown
below.