PEL-3000 User Manual

112

Data Edit
Overview

Each step in a fast sequence contains the
following setting parameters:

Setting

Setting Range

Description

Step

0001 ~ 1000

Selects/displays the current
step in the sequence.

The number of available
steps is dependent on the
number of steps added using
the Ins. Point[F1] functions.

A minimum of 3 steps.

Value

The current or resistance
setting for the selected
operating mode.

TRIG OUT

ON, OFF

When TRIG OUT is set to
ON, a trigger signal is
output from the TRIG OUT
BNC terminal at the start of
the step. See page 150 for
details.

Time

amplitude

TRIG OUT = ON

TRIG OUT

Start of step

FILL Overview

The FILL function is used to evenly step up the
current or resistance value settings from a
starting step to a finishing step.

The Fill Function can be used before or after
points are added to the fast sequence.

Before: Will pre-fill each value within the fill
range when a new step is added.

After: Will post-fill each value within the fill