beautypg.com

Monitor signal output – GW Instek PEL-3000 Series User Manual User Manual

Page 152

background image

PEL-3000 User Manual

150

Monitor Signal Output

Trigger Signal Output

Description

The trigger output signal is generated every
time a switching operation is performed (i.e.,
Dynamic mode) or when a Fast or Normal
Sequence is executed and the TRIG OUT
parameter is enabled.

The trigger output signal from TRIG OUT BNC
is a 5V pulse of at least 2us with an impedance
of 500Ω. The common potential is connected to
the chassis potential. The signal threshold level
is TTL.

Time

amplitude

TRIG OUT = ON

TRIG OUT

Start of step
or switch
operation

Current Monitor Output

Description

The voltage output from the IMON OUT
terminal and from the IMON pin on the J1
connector is used to represent the current input
level.

The voltage range used to represent the full
scale current range from the IMON OUT
terminal and from the IMON pin on the J1
connector depends on the current range