Shorting the load – GW Instek PEL-3000 Series User Manual User Manual

Page 62

background image

PEL-3000 User Manual

60

Note

The load can be set to automatically turn on at
start up. See page 75.

The load can be turned on via remote control.
See the programming manual.

The load can be turned on via external control.
See page 143.

By default the load will automatically turn off if
the range or operating mode (CC, CV, CR, CP) is
changed. To disable this behavior, Set Load Off
(Mode)
and Load Off (Range) to the OFF setting.
See page 76 for details.

Display

Mode

CP

IRange

H 35A

VRange

L 15V

Function

Dynamic

Configure

1.500

V

0.000

A

1.50

w

Timer1

0.025

Level2

0.00

Level1

0.00

LOAD

RS232

Fine

03/Sep/2012

mS

W

W

LOAD on

Shorting the Load

Description

The Short key can be used to simulate a short
circuit of the load input terminals. A short
circuit is simulated by:

Setting the current to the maximum value in CC
mode.

Setting the resistance to the minimum value in
CR mode.

Setting the voltage to the minimum value in CV
mode.

Setting the power to the maximum value in CP
mode.

When the load is shorted, the external controller
also sends a short signal. See page 149 for usage
details.