beautypg.com

Cr mode, Cv mode – GW Instek PEL-3000 Series User Manual User Manual

Page 199

background image

APPENDIX

197

CR Mode

Model

PEL-3021

PEL-3041

PEL-3111

Operating Range

*1

H Range

23.3336S~400uS
(42.857mΩ~2.5kΩ)

46.6672S~800uS
(21.428mΩ~1.25kΩ)

140.0016S~2.4mS
(7.1427mΩ
~416.6667Ω)

M Range

2.33336S~40uS
(428.566mΩ~25kΩ)

4.6667S~80uS
(214.28mΩ~12.5kΩ)

14.0001S~242.4uS
(71.427mΩ~
4.16667kΩ)

L Range

0.233336S~4uS
(4.28566Ω~250kΩ)

0.46667S~8uS
(2.1428Ω~125kΩ)

1.40001S~24.24uS
(714.27mΩ
~41.6667kΩ)

Setting Range
H Range

24.5S~0S
(40.8163 mΩ~OPEN)

49.0S~0 S
(20.408 mΩ~OPEN)

147.000S~0S
(6.8027 mΩ~OPEN)

M Range

2.45S~0S
(408.1633mΩ~
OPEN)

4.90S~0S
(204.08mΩ~OPEN)

14.70000S~0S
(68.0272mΩ~OPEN)

L Range

0.245S~0S
(4.08163Ω~OPEN)

0.490S~0S
(2.0408Ω~OPEN)

1.4000S~0S
(680.2721mΩ~OPEN)

Resolution
H Range

400uS

800uS

2.4mS

M Range

40uS

80uS

240uS

L Range

4uS

8uS

24uS

Accuracy of Setting

*2

H, M Range

±(0.5 % of set

*3

+ 0.5 % of f.s.

*4

) + Vin

*5

/500 kΩ

L Range

±(0.5 % of set

*3

+ 0.5 % of f.s.) + Vin

*5

/500 kΩ


*1 Siemens[S] = Input current[A] / Input voltage[V] = 1 / resistance[Ω]
*2 Converted value at the input current. At the input current. It is not applied for

the condition of the parallel operation.

*3 set = Vin / Rset
*4 f.s. = Full scale of High Range
*5 Vin = Input terminal voltage of electronic load

CV Mode

Model

PEL-3021

PEL-3041

PEL-3111

Operating Range
H Range

1.5V~150V

1.5V~150V

1.5V~150V

M Range

1.5V~15V

1.5V~15V

1.5V~15V