PEL-3000 User Manual

82

Display

2.596

V

500.06

mA

1.2982

w

CV B Value

15.000

CV A Value

15.000

LOAD

RS232

A Value

Fine

03/Sep/2012

V

V

OPP Alarm

Please Press Enter

To Clear Alarm

OPP

OCP indicator

Alarm message
when OPP is set
to Load Off

Mode

CV

I Range

L 0.7A

V Range

L 15V

Response

Fast

Configure

UVP

Description

If the UVP is tripped, the PEL-3000 will turn off
the load.

The UVP levels can be set from 0V to 10%
higher than the rating voltage.

Operation

1. Press

Main

> Configure[F5] > Protection[F1]

and set the UVP Level.

Range:

UVP Level: OFF, 0~Rating voltage +
10%

Alarm

The UVP indicator and a message will only
appear on the screen when the input voltage is
below the UVP level. The Enter key must be
pressed to clear the alarm message.

To clear the UVP indicator, remove the cause of
the under voltage ~ i.e., increase the input
voltage.