5. Use the selector keys and

scroll wheel or number pad
to enter the FSK rate.

/

6. Press F1~F5 to select the

frequency unit.

~

Range FSK

Rate 2mHz~100kHz

Default 10Hz

Note

If an external source is selected, FSK Rate settings are
ignored.

FSK Source

The AFG-3000 accepts internal and external FSK sources, with
internal as the default source. When the FSK source is set to internal,
the FSK rate is configured using the FSK Rate function. When an
external source is selected the FSK rate is equal to the frequency of
the Trigger INPUT signal on the rear panel.

Panel Operation

1. Press the MOD key.

2. Press F3 (FSK).