beautypg.com

Source[1]:am:internal:function, Source[1]:am:internal:frequency – GW Instek AFG-3000 Series User Manual

Page 225

background image

REMOTE

INTERFACE

225

Sets the modulation source to external.

Query Syntax

SOURce[1]:AM:SOURce?

Return Parameter INT Internal

EXT External

Example

SOUR1:AM:SOUR?
INT

The modulation source is set to internal.

SOURce[1]:AM:INTernal:FUNCtion

Source Specific
Command

Description

Sets the shape of the modulating waveform from
sine, square, triangle, upramp and dnramp. The
default shape is sine.

Note

Square and triangle waveforms have a 50% duty
cycle. Upramp and dnramp have a symmetry of
100% and 0%, respectively.

Syntax

SOURce[1]:AM:INTernal:FUNCtion
{SINusoid|SQUare|TRIangle|UPRamp|DNRamp}

Example

SOUR1:AM:INT:FUNC SIN

Sets the AM modulating wave shape to sine.

Query Syntax

SOURce[1]:AM:INTernal:FUNCtion?

Return Parameter SIN Sine UPRAMP Upramp

SQU Square

DNRAMP Dnramp

TRI Triangle

Example

SOUR1:AM:INT:FUNC?
SIN

The shape for the modulating waveform is Sine.

SOURce[1]:AM:INTernal:FREQuency

Source Specific
Command

Description

Sets the frequency of the internal modulating
waveform only. The default frequency is 100Hz.