REMOTE

INTERFACE

243

SOURce[1]:FREQuency:STARt

Source Specific
Command

Description

Sets the start frequency of the sweep. 100Hz is the
default start frequency.

Note

To sweep up or down, set the stop frequency
higher or lower than the start frequency.

Syntax

SOURce[1]:FREQuency:STARt
{|MINimum|MAXimum}

Parameter

100μHz~ 80MHz(3081)/
50MHz(3051)
100μHz~ 1MHz (Ramp)

Example

SOUR1:FREQ:STAR +2.0000E+03

Sets the start frequency to 2kHz.

Query Syntax

SOURce[1]:FREQuency:STARt? [MINimum|
MAXimum]

Return Parameter

Returns the start frequency
in Hz.

Example

SOUR1:FREQ:STAR? MAX
+8.0000E+07

Returns the maximum start frequency allowed.

SOURce[1]:FREQuency:STOP

Source Specific
Command

Description

Sets the stop frequency of the sweep. 1 kHz is the
default start frequency.

Note

To sweep up or down, set the stop frequency
higher or lower than the start frequency.

Syntax

SOURce[1]:FREQuency:STOP
{|MINimum|MAXimum}

Parameter

100μHz~ 80MHz(3081)/
50MHz(3051)
100μHz~ 1MHz (Ramp)