beautypg.com

Source[1]:apply:noise, Source[1]:apply:triangle – GW Instek AFG-3000 Series User Manual

Page 208

background image

AFG-3000 Series User Manual

208

Parameter

500μHz~50MHz

<

amplitude

>

10mV~10V (50Ω)

0~4.99V (50Ω)

Example

SOUR1:APPL:PULS 1KHZ,MIN,MAX

Sets frequency to 1kHz and sets the amplitude to
minimum and the and offset to the maximum.

SOURce[1]:APPLy:NOISe

Source Specific
Command

Description

Outputs Gaussian noise with a 50 MHz
bandwidth. Amplitude and offset can also be set.

Note

Frequency cannot be used with the noise function;
however a value (or DEFault) must be specified.
The frequency is remembered for the next function
used.

Syntax

SOURce[1]:APPLy:NOISe [
[, [,] ]]

Parameter

Not applicable

<

amplitude

>

10mV~10V (50Ω)

0~4.99V (50Ω)

Example

SOUR1:APPL:NOIS DEF, 3.0, 1.0

Sets the amplitude to 3 volts with an offset of 1
volt.

SOURce[1]:APPLy:TRIangle

Source Specific
Command

Description

Outputs a triangle wave from the selected channel
when the command has executed. Frequency,
amplitude and offset can also be set.

Syntax

SOURce[1]:APPLy:TRIangle [
[, [,] ]]

Parameter

1μHz~1MHz

<

amplitude

>

10mV~10V (50Ω)