-313 Calibration memory lost;memory corruption detected
Indicates that a fault (check sum error) has occurred
with the non-volatile memory that stores the
calibration data.
-314 Save/recall memory lost;memory corruption detected
Indicates that a fault (check sum error) has occurred
with the non-volatile memory that stores the
save/recall files.
-315 Configuration memory lost;memory corruption detected
Indicates that a fault (check sum error) has occurred
with the non-volatile memory that stores the
configuration settings.
-350 Queue overflow
Indicates that the error queue is full (over 20
messages generated, and not yet read). No more
messages will be stored until the queue is empty.
The queue can be cleared by reading each message,
using the *CLS command or restarting the function
generator.
-361 Parity error in program message
Indicates that there is a RS232 parity setting
mismatch between the host PC and the function
generator.
-362 Framing error in program message
Indicates that there is a RS232 stop bit setting
mismatch between the host PC and the function
generator.
-363 Input buffer overrun
Indicates that too many characters have been sent to
the function generator via RS232. Ensure
handshaking is used.