26

Selecting a Waveform

Square Wave

Example: Square Wave, 3Vpp, 75%Duty, 1 kHz

Output

1. Press the Waveform

key and select
Square (F2).

2. Press Duty(F1),

followed by 7 + 5 +
%(F5)

Input: N/A

3. Press the Freq/Rate

key, followed by 1 +
kHz (F5).

4. Press the AMPL

key, followed by 3 +
VPP (F6).

5. Press the output

key.

Triangle Wave

Example: Triangle Wave, 5Vpp,10kHz

Output

Input: N/A

1. Press the Waveform

key and select
Triangle (F3).

2. Press the Freq/Rate

key, followed by 1 +
0 + kHz (F5).