beautypg.com

Infinite burst count – GW Instek AFG-3000 Series User Manual

Page 112

background image

AFG-3000 Series User Manual

112

5. Use the selector keys and

scroll wheel or number pad
to enter the number of
cycles.

/

6. Press F5 to select the Cyc

unit.

Range Cycles

1~1,000,000

Note

Burst cycles are continuously output when the
internal trigger is selected. The burst period
determines the rate of bursts and the time between
bursts.

Burst cycle must be less than the product of the
burst period and wave frequency.

Burst Cycle < (Burst Period x Wave Frequency)

If the burst cycle exceeds the above conditions, the
burst period will be automatically increased to
satisfy the above conditions.

If gated burst mode is selected, burst cycle is
ignored. Though, if the burst cycle is changed
remotely whilst in gated mode, the new burst cycle
is remembered when used next.

Infinite Burst Count

Panel Operation

1. Press the Burst key.

2. Press F1 (N Cycle).