beautypg.com

Burst frequency – GW Instek AFG-3000 Series User Manual

Page 110

background image

AFG-3000 Series User Manual

110

In Gated mode, burst count, burst cycle and trigger source are
ignored. If a trigger is input, then the trigger will be ignored and will
not generate any errors.

Panel Operation

1. Press the Burst key.

2. Select either N Cycle (F1) or

Gate (F2).

~

Burst Frequency

In the N Cycle and Gated modes, the waveform frequency sets the
repetition rate of the burst waveforms. In N-Cycle mode, the burst is
output at the waveform frequency for the number of cycles set. In
Gated mode the waveform frequency is output while the trigger is
high. Burst mode supports sine, square, triangle or ramp
waveforms.

Panel Operation

1. Press the FREQ/Rate key.

2. The FREQ parameter will become highlighted

in the parameter window.

3. Use the selector keys and

scroll wheel or number pad
to enter the frequency.

/

4. Press F2~F6 to choose the

frequency unit.

~

Range Frequency 2mHz~80MHz(3081)/

50MHz(3051)