beautypg.com

Source[1]:sweep:source, Output[1]:trigger:slope – GW Instek AFG-3000 Series User Manual

Page 247

background image

REMOTE

INTERFACE

247

SOURce[1]:SWEep:SOURce

Source Specific
Command

Description

Sets or queries the trigger source as immediate
(internal), external or manual. Immediate (internal)
is the default trigger source. IMMediate will
constantly output a swept waveform. EXTernal
will output a swept waveform after each external
trigger pulse. Manual will ouput a swept
waveform after the trigger softkey is pressed.

Note

If the APPLy command was used to create the
waveform shape, the source is automatically set to
IMMediate.
The *OPC/*OPC? command/query can be used to
signal the end of the sweep.

Syntax

SOURce[1]: SWEep:SOURce {IMMediate|EXTernal|
MANual}

Example

SOUR1: SWE:SOUR EXT

Sets the sweep source to external.

Query Syntax

SOURce[1]: SWEep:SOURce?

Return Parameter IMM Immediate

EXT External

MANual Manual

Example

SOUR1:SWE:SOUR?
IMM

The sweep source is set to immediate.

OUTPut[1]:TRIGger:SLOPe

Source Specific
Command

Description

Configures the trigger output signal (TTL) as a
positive or negative slope. A positive slope will
output a pulse with a rising edge and a negative
slope will output a pulse with a falling edge.