REMOTE

INTERFACE

209

0~4.99V (50Ω)

Example

SOUR1:APPL:TRI 2khz, 3.0, 1.0

Sets the frequency to 1 MHz with an amplitude of
3 volts and with an offset of 1 volt.

SOURce[1]:APPLy:DC

Source Specific
Command

Description

Outputs a triangle wave from the selected channel
when the command has executed. Frequency,
amplitude and offset can also be set.

Note

Frequency and amplitude cannot be used with the
DC function; however a value (or DEFault) must
be specified. The values are remembered for the
next function used.

Syntax

SOUR[1]:APPLy:DC [
[,|DEFault> [,] ]]

Parameter

Not applicable

<

amplitude

> Not

applicable

±5V (50Ω), ±10V (open)

Example

SOUR1:APPL:DC DEF, DEF, 1.0

Sets the DC offset to 1 volt.

SOURce[1]:APPLy:USER

Source Specific
Command

Description

Outputs an arbitrary waveform from the selected
channel. The output is that specified from the
FUNC:USER command.

Note

Frequency and amplitude cannot be used with the
DC function; however a value (or DEFault) must
be specified. The values are remembered for the
next function used.

Syntax

SOURce[1]:APPLy:USER [ [,
[,] ]]