beautypg.com

Source[1]:fm:state, Source[1]:fm:source – GW Instek AFG-3000 Series User Manual

Page 229

background image

REMOTE

INTERFACE

229

SOURce[1]:FM:STATe

Source Specific
Command

Description

Sets or disables FM modulation. By default FM
modulation is disabled. FM modulation must be
enabled before setting other parameters.

Note

Burst or sweep mode will be disabled if FM
modulation is enabled. As only one modulation is
allowed at any one time, other modulation modes
will be disabled when FM modulation is enabled.

Syntax

SOUR[1]:FM:STATe {OFF|ON}

Example

SOUR1:FM:STAT ON

Enables FM modulation.

Query Syntax

SOURce[1]:FM:STATe?

Return Parameter 0 Disabled

(OFF)

1 Enabled

(ON)

Example

SOUR1:FM:STAT?
1

FM modulation mode is currently enabled.

SOURce[1]:FM:SOURce

Source Specific
Command

Description

Sets or queries the modulation source as internal
or external. Internal is the default modulation
source.

Note

If an external modulation source is selected,
modulation depth is limited to ± 5V from the MOD
INPUT terminal on the rear panel. For example, if
modulation depth is set to 100%, then the
maximum amplitude is +5V, and the minimum
amplitude is -5V.

Syntax

SOURce[1]:FM:SOURce {INTernal|EXTernal}