-221 Settings conflict;offset changed due to amplitude

Example: The offset is not a valid offset value, it is
automatically adjusted, considering the amplitude.
|offset|≤ max amplitude – Vpp/2

-221 Settings conflict;amplitude changed due to offset

Example: The amplitude is not a valid value, it is
automatically adjusted, considering the offset.
Vpp ≤ 2X (max amplitude -|offset|)

-221 Settings conflict;low level changed due to high level

Example: The low level value was set too high. The
low level is set 1 mV less than the high level.

-221 Settings conflict;high level changed due to low level

Example: The high level value was set too low. The
high level is set 1 mV greater than the low level.

-222 Data out of range;value clipped to upper limit

Example: The parameter was set out of range. The
parameter is automatically set to the maximum
value allowed.
SOURce[1]:FREQuency 80.1MHz.

-222 Data out of range;value clipped to lower limit

Example: The parameter was set out of range. The
parameter is automatically set to the minimum value
allowed.
SOURce[1]:FREQuency 0.1μHz.

-222 Data out of range;period; value clipped to ...

Example: If the period was set to a value out of
range, it is automatically set to an upper or lower
limit.