beautypg.com

Source[1]:frequency:center – GW Instek AFG-3000 Series User Manual

Page 244

background image

AFG-3000 Series User Manual

244

Example

SOUR1:FREQ:STOP +2.0000E+03

Sets the stop frequency to 2kHz.

Query Syntax

SOURce[1]:FREQuency:STOP? [MINimum|
MAXimum]

Return Parameter

Returns the stop frequency
in Hz.

Example

SOUR1:FREQ:STOP? MAX
+8.0000E+07

Returns the maximum stop frequency allowed.

SOURce[1]:FREQuency:CENTer

Source Specific
Command

Description

Sets and queries the center frequency of the sweep.
550 Hz is the default center frequency.

Note

The maximum center frequency depends on the
sweep span and maximum frequency:
max center freq = max freq – span/2

Syntax

SOURce[1]:FREQuency:CENTer
{|MINimum|MAXimum}

Parameter

100μHz~ 80MHz(3081)/
50MHz(3051)
100μHz~ 1MHz (Ramp)

Example

SOUR1:FREQ:CENT +2.0000E+03

Sets the center frequency to 2kHz.

Query Syntax

SOURce[1]:FREQuency:CENTer? [MINimum|
MAXimum]

Return Parameter

Returns the stop frequency
in Hz.

Example

SOUR1:FREQ:CENT? MAX
+8.0000E+06

Returns the maximum center frequency allowed,
depending on the span.