Calibration correction – Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 24
3-8
M-4659
Low-cut (Can be set when the mode is 1-5V, and when the mode is VOLT with
square root (SQRT) selected. )
Select [ON] to use the low-cut function.
Low-cut value (Can be set when the mode is VOLT with square root (SQRT)
selected.)
Set the low-cut value in the range of 0.0% to 5.0% of the input span.
Input value
Low-cut value
Result of square root computation
Measured value
Calibration Correction
Set the input and output values for the calibration correction. The number of set points
(including the start and end points) can be specified in the range 2 to 16.
Click to delete the selected row.
Click to add set points (rows) to the number of calibration set points.
Input
Number of set points
Calibration set points
Y5
Y4
Y3
Y2
Y1
Output
X1 X2 X3 X4
X5
Condition
X1 < X2 < X3 < X4 < X5
Selectable Range of Input and Output Values
• Channels on which linear scaling is specified
–30000 to 30000 (the decimal place is the same setting as the scale value)
• Other channels
Value in the measurable range of the selected range
Example: –2.0000 to 2.0000 for 2 V range
3.3 Setting the Measurement Channels, Ext. Channels