Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 142
6-27
M-4659
Configuring the MV1000/MV2000
1
2
3
4
5
6
7
8
Index
File
Directory name
Set the name of the directory on the storage medium for saving the data on the external
storage medium. (Up to 20 characters)
Symbols that can be used: #, %, (, ), +, -, ., @, °, and _.
Strings that cannot be used: AUX, CON, PRN, NUL, CLOCK, COM1 to COM9, and LPT1
to LPT9.
Header
Set the header comment to be written to the data file. (Up to 50 characters)
Structure
Sets the structure of the file name when saving data.
Settings
Description
Date
Serial number + user-assigned character string + date
Serial
Serial number + user-assigned character string
Batch
Serial number + batch name (when using the batch function)
File name
Set the user-assigned section of the file name. (Up to 16 characters)
Symbols that can be used: #, %, (, ), +, -, ., @, °, and _.
File Format
Settings
Description
Text
Display data files and event data files are in text format.
Binary
Display data files and event data files are in binary format.
Field Title, Field Characters
Set the string.
Title of field: Up to 20 characters, Characters: Up to 30 characters
6.5 Entering General Settings