Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 125

background image

6-10

M-4659

Moving Average

To use the moving average, select the sampling count [Times] (2 to 400).

Tag

Up to 16 characters can be entered for the tag.
You can use the tag name instead of the channel number to be displayed on the screen.
This can be selected when [Tag] is [Tag] under [Detail Setting] in the [Basic Setting] tab.

Memory Sampling

Turn [ON] (sample) or [OFF] (do not sample).

Display Zone (Zone L and U)

You can select the range of the screen in which the waveform of each channel is to be
displayed.
Specify positions (%) on the display scale for the upper and lower limits.
The conditions for setting the zones are as follows:
• Range: 0% to 100%
The lower limit L must be less than the upper limit
• The difference between the lower and upper limits is at least 5%.

Graph

For details, see section 5.7 in the RD-MV1000/RD-MV2000 User’s Manual (IM RD-
MV1000-01E).

Scale display position

Select the scale display position on the trend display from 1 to 10 for the RD-MV2000 or
from 1 to 6 for the RD-MV1000. Select [OFF] if you do not wish to display the scale.

Scale divide position

Select the number of main scale marks on the trend display from 4 to 12 and C10.
C10: The scale is equally divided into 10 sections by main scale marks, and scale

values are indicated at 0, 30, 50, 70, and 100% positions on the trend display.

Bar display position

Select [Normal], [Center], [Lower], or [Upper].

Bar divide number

Select number of divisions of the scale on the bar graph display.

6.3 Setting the Measurement Channels, Ext. Channels